Common Faults and Troubleshooting Methods for Helium Mass Spectrometer Leak Detectors

Publishdatea:2019-09-26 Views:128

Based on feedback from NOY customers on the helium mass spectrometer leak detector, the following table provides a reference for common faults and troubleshooting methods with a high failure rate during operation:

Category

Fault Symptom

Inspection and Analysis Method

Power Supply

Leak detector does not run after power-on

Check whether the power outlet has AC 220V;

Check whether the power connection is secure;

Check whether the power switch outputs 220V;

Check whether the fuse is blown

No display after power-on

Check whether the display module cable connection is loose

Mechanical Pump

Display is on after power-on, but the mechanical pump does not run

Check whether the mechanical pump switch is turned on;

Check whether the mechanical pump power cable is securely connected

Abnormal noise from the mechanical pump

Check all connection clamps on the mechanical pump;

Check whether the leak test port is sealed;

Check whether the mechanical pump fan guard is deformed;

Mechanical pump not OK

Check for large leaks;

Check whether the mechanical pump oil has solidified

Ion Source

Filament does not energize

Check whether the molecular pump has completed startup;

Check whether the ion source socket is loose

Ion source repeatedly energizes

Check the wiring sequence of the ion source socket for incorrect connection

Based on feedback from NOY customers on the helium mass spectrometer leak detector, the following table provides a reference for common faults and troubleshooting methods with a high failure rate during operation:

Category

Fault Symptom

Inspection and Analysis Method

Power Supply

Leak detector does not run after power-on

Check whether the power outlet has AC 220V;

Check whether the power connection is secure;

Check whether the power switch outputs 220V;

Check whether the fuse is blown

No display after power-on

Check whether the display module cable connection is loose

Mechanical Pump

Display is on after power-on, but the mechanical pump does not run

Check whether the mechanical pump switch is turned on;

Check whether the mechanical pump power cable is securely connected

Abnormal noise from the mechanical pump

Check all connection clamps on the mechanical pump;

Check whether the leak test port is sealed;

Check whether the mechanical pump fan guard is deformed;

Mechanical pump not OK

Check for large leaks;

Check whether the mechanical pump oil has solidified

Ion Source

Filament does not energize

Check whether the molecular pump has completed startup;

Check whether the ion source socket is loose

Ion source repeatedly energizes

Check the wiring sequence of the ion source socket for incorrect connection